Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Health Products, Inc.: In vitro Test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Reagent Pack 100 test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Calibrators are composed of six vials of calibrators A, B, C, D, E, and F containing cyclosporine prepared in human blood component (non-reactive for HBxAg, anti-HCv, anti-HIV-1/HIV-20. TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Controls are composed of three vials L, M and H containing cyclosporine in human blood componenet (nonreactive for HBsAg, anti-HCV, and anti-HIV-1/HIV-2).

Recall number
Z-0624-06
Recalling firm
Abbott Health Products, Inc.
Product
In vitro Test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Reagent Pack 100 test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Calibrators are composed of six vials of calibrators A, B, C, D, E, and F containing cyclosporine prepared in human blood component (non-reactive for HBxAg, anti-HCv, anti-HIV-1/HIV-20. TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Controls are composed of three vials L, M and H containing cyclosporine in human blood componenet (nonreactive for HBsAg, anti-HCV, and anti-HIV-1/HIV-2).
Status
Terminated
Initiated
2005-03-23
Posted
2006-03-15
Terminated
2012-01-27
Reason
Patient results reported using calibrators made from new primary material may be falsely decreased by 5-8% compared to results using calibrators made from the previous primary material. An Important Product Information Letter with technical information regarding the Whole Blood Calibrators and Controls, implementing new internal reference standards was issued.
Root cause
Other
Action
Abbott has recently implemented new internal reference standards for TDx/TDxFLx Cyclosporine monoclonal Whole Blood Calibrators and Controls. A change in performance is associated with the use of calibrator lots 1829Q100 and higher and control lots 18230Q100 and higher, which are the first lots manufactured using the new internal reference standards. The specific performance changes are identified in the letter, dated March 23, 2005, as compared to current product using the previous standards (calibrators lot 05434Q100 and control lot 03071Q100).
Quantity
4,514 kits were distributed
Distribution
Nationwide, Puerto Rico, Hawaii, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Chile, Uruguay, Argentina, Brazil, Dominican Republic, Canada, Germany, Japan , Singapore, India, Taiwan, Thailand, Korea, Australia, Honduras, and Panama,
Lot, serial or model codes
Lot 18229Q100 (exp 8/26/2006) Lot 18230Q100 (exp 10/05/2006)
510(k) numbers
not on file
PMA numbers
["P890025"]
Product code
MGU
Firm FEI
2623532
Firm city
Barceloneta
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register