Greatbatch Medical: Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration.

Recall number
Z-0623-2017
Recalling firm
Greatbatch Medical
Product
Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration.
Status
Terminated
Initiated
2016-09-08
Posted
2016-11-18
Terminated
2017-09-08
Reason
Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.
Root cause
Process design
Action
Greatbatch sent a "Urgent Medical Device Recall" letter dated September 7, 2016 to all affected customers .The letter described the product, "Reason for Field Correction Action", "Risk to Health" , "Replacement Devices", & "Instructions to Customers" . Advised customers to quarantine the product and contact end users and share the recall information with them. Requested consignees to complete & return the Field Recall Response Form. For questions contact your local Greatbatch Medical representative, contact us at 1-763-951-8235, or e-mail us at FieldActionCenter@Greatbatch.com.
Quantity
23,075
Distribution
US: NY, TX.
Lot, serial or model codes
GBP3176925, GBP3176927, VN3470927, VN3470929, VN3486000, VN3485998, VN3495047, GBP3421664, VN3433259, VN3451654, VN3487654, VN3491069, VN3470925, VN3451646.
510(k) numbers
["K071574"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2183787
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register