B. Braun Medical, Inc.: Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a pole.

Recall number
Z-0623-2013
Recalling firm
B. Braun Medical, Inc.
Product
Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a pole.
Status
Terminated
Initiated
2012-03-23
Posted
2012-12-31
Terminated
2015-02-02
Reason
Firm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps.
Root cause
Device Design
Action
B. Braun Medical, Inc sent an Urgent Medical Device letter dated March 23, 2012 , to all affected customers. The letter identified the product the problem and the action needed by the customer to take. B. Braun has implemented an improvement to the pump locking mechanism of t he older style pole clamps and will be upgrading any older style pole clamps that may remain within your facility. Please return the completed form from attachment 1 in the self-addressed envelope provided or fax the form to (610) 849-1197 within two (2) weeks of receipt\. If you require any additional information or have any questions, comments or concerns please contact your B. Braun Medical Customer Service at 800-627-7867.
Quantity
13,078
Distribution
Worldwide Distribution--US (nationwide) including the states of AL, AZ, NC and TN.
Lot, serial or model codes
Model number 8713130 and lot numbers 209, 309, 409, 509, 809, 909, 60939747, 60943679, 60945966, 60948966, 60959578, 60960830, 60962412, 61170378, 5048020067, 5048805107, 5048920117, 5049152127, 5049484028, 5049545028, 5049614028, 5049704038, 5049788048, 5049821048, 5050123058, 5050297068, 5050385078, 5051802039, 5051879039, 5051893059, 5051940049, 5051999049, 5052519099, 5052638099, 5052690109, 5052692109, 5052711109, 5052726119, 5052788119, 5052920119, 5053045129, 5053112010, 5053129129, 5053146010, 5053207010, 5053269020, 5053449020, 5053505030, 5053534040, 5053584030, 5053605040, and 5053645040.
510(k) numbers
["K062700","K083689"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2523676
Firm city
Allentown
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register