MAQUET Inc.: Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in the 1009.8 1 series are accessory components which can be connected to certain MAQUET operating tables. Several of the table top adjustment motions (depending on the table and foot switch versions in use) can then be executed by the foot. The foot switch is equipped with rocker switches. The switching elements are used to initiate the table top functions.

Recall number
Z-0620-2009
Recalling firm
MAQUET Inc.
Product
Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in the 1009.8 1 series are accessory components which can be connected to certain MAQUET operating tables. Several of the table top adjustment motions (depending on the table and foot switch versions in use) can then be executed by the foot. The foot switch is equipped with rocker switches. The switching elements are used to initiate the table top functions.
Status
Terminated
Initiated
2008-12-08
Posted
2009-01-15
Terminated
2012-06-18
Reason
A bearing point of the pedal of the foot switch may come loose. With continued use of the foot switch, one or both springs beneath the pedal can fall out of their mounting, and it is possible the pedal comes to rest on the switch causing an unintended motion of the operating table during patient use.
Root cause
Device Design
Action
An Urgent Device Correction letter (with an enclosed Customer Response form) was sent via Federal Express on Monday, December 8, 2008 to all customers who had received tables. Customers were notified in the letter of the problem and actions to be taken including not to use any affected foot pedal for safety reasons. In addition, customers were instructed to complete and return the Customer Response Form via fax (732-667-1903) to Maquet Inc. Please call USA Maquet, Inc. at toll free number 1-888-627-8383 #1 (then select the option for Surgical Workplace products) for questions.
Quantity
587 units total (54 units in US)
Distribution
Worldwide Distribution.
Lot, serial or model codes
Product manufactued between 8/2004 and 9/2008 may be affected: 1009.81A0, 1009.81A1, 1009.81A2, 1009.81A3, 1009.81B0, 1009.81B1, 1009.81B2, 1009.81C0, 1009.81C1, 1009.81D0, 1009.81D1, 1009.81D2, 1009.81D3, 1009.81D4, 1009.81E0, 1009.81E1, 1009.81E2, 1009.81E3, 1009.81F0, 1009.81F1, 1009.81F2, 1009.81F3, 1009.81G0, 1009.81G1 and 1009.81G2.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FWZ
Firm FEI
3008355164
Firm city
Bridgewater
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register