FDA Medical Device Recalls (openFDA)
bioMerieux, Inc.: VITEK 2 Antimicrobial Susceptibility Testing (AST) using one or more of these susceptibility products: Catalog # 410028, VITEK 2 AST-ST01 Catalog # 420915, VITEK 2 AST-ST02 Catalog # 421040, VITEK 2 AST-ST03 The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of S. pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
- Recall number
- Z-0618-2019
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK 2 Antimicrobial Susceptibility Testing (AST) using one or more of these susceptibility products: Catalog # 410028, VITEK 2 AST-ST01 Catalog # 420915, VITEK 2 AST-ST02 Catalog # 421040, VITEK 2 AST-ST03 The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of S. pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
- Status
- Terminated
- Initiated
- 2018-10-22
- Posted
- not on file
- Terminated
- 2019-11-04
- Reason
- False resistant results for Streptococcus anginosus and Streptococcus constellatus strains were reported.
- Root cause
- Component change control
- Action
- The Firm, Biomerieux, issued a " URGENT Product Correction Notice" to their consignees informing of the matter and providing certain actions for them to take. Each Notice will be accompanied with an Acknowledge Form which is to be returned to bioMerieux.. The consignees were instructed to take the following action at this time: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " Reference your Laboratory Standard Operating Procedures to identify card types and breakpoints used in your facility to determine if your laboratory is impacted by the issue described in this letter. " Reference your Laboratory Standard Operating Procedures to identify if retrospective analysis is required for reported results associated with the referenced limitation. " For systems utilizing EUCAST breakpoints and operating with VITEK 2 Systems Software version 8.01, create a bioART Rule for the referenced limitation associated with Ceftriaxone (cro01n). " Please store this letter with your bioMrieux instrument documentation. -Please complete the attached Acknowledgement Form (Attachment A) and return it to bioMrieux, Inc. If you have any questions or concerns, please contact your local bioMrieux representative or call , Regulatory Quality Compliance Specialist, at 314-731-8805 or email: debra.broyles@biomerieux.com.
- Quantity
- 184,374
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY including Puerto Rico and to countries of:Algeria, Angola, Azerbaijan, Bahrain, Bangladesh, Belarus, Sarl, Benin, Bosnia-Herz, Cameroon, Chad, Cote d'Ivoire, Croatia, Dem. Rep. Congo, Egypt, , Georgia, Guam, Ethiopia, Iraq, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, UAB, Lithuania, Macedonia, Malta, Mongolia, Morocco, Mozambique, Niger, Oman, Pakistan, Palestine, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovenia, Sri Lanka, Tunisia, Ukraine, Gabon, and Zimbabwe.
- Lot, serial or model codes
- Catalog # 410028, VITEK 2 AST-ST01 Catalog # 420915, VITEK 2 AST-ST02 Catalog # 421040, VITEK 2 AST-ST03
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LON
- Firm FEI
- 3002769706
- Firm city
- Durham
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28