Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Endoscopy: Stryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico

Recall number
Z-0618-2011
Recalling firm
Stryker Endoscopy
Product
Stryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico
Status
Terminated
Initiated
2009-08-20
Posted
2010-12-14
Terminated
2010-12-14
Reason
Potential breakage -- product may break while inserting it into the bone during surgery. This may prolong surgery, or, if undetected during surgery, result in retained fragments that may cause metal sensitization or immune response, or necessitate further intervention to retrieve the titanium fragments.
Root cause
Material/Component Contamination
Action
Stryker issued an Urgent Device Removal notification letter dated August 21, 2009 to consignees. Consignees were asked to acknowledge receipt and distribute the letter, track reponses, and destroy any returned material and document through weekly summary and records. Stryker can be contacted at 408 754-2000 about this recall.
Quantity
447 units
Distribution
Nationwide Distribution: to one distributor in NJ.
Lot, serial or model codes
Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, serial numbers 09049AE2, 09050AE2, 09051AE2, 09052AE2, 09054AE4
510(k) numbers
["K071157"]
PMA numbers
not on file
Product code
HWC
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register