Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road, Hoffman Estates, IL 60195-5203

Recall number
Z-0617-03
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road, Hoffman Estates, IL 60195-5203
Status
Terminated
Initiated
2003-02-10
Posted
2003-03-06
Terminated
2004-04-30
Reason
Unintended yoke rotational motion.
Root cause
Other
Action
Siemens sent Customer Advisory letters dated 2/10/03 to all Integrated Orbiter customers. The accounts were informed of the problem with unexpected camera movement in the opposite direction at the beginning of a SPECT study, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended camera rotation. The upgrade will be done free of charge.
Quantity
256 units
Distribution
Nationwide, United Arab Emirates, Argentina, Austria, Azerbaizan,Bangladesh, Belgium, Brazil, Canada, China, Colombia, Cyprus, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Jordan, Japan, Korea, Norway, Pakistan, Romania, Switzerland, Tunisia, Venezuela and the Netherlands.
Lot, serial or model codes
model 35-26-932, serial numbers 102, 1001/1217, 1169A; model 35-44-637, serial numbers 0101/0142, 120A
510(k) numbers
["K850397"]
PMA numbers
not on file
Product code
IYX
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register