Skip to the content

FDA Medical Device Recalls (openFDA)

Landauer: Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Recall number
Z-0616-2024
Recalling firm
Landauer
Product
Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Status
Completed
Initiated
2023-11-15
Posted
2023-12-22
Terminated
not on file
Reason
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
Root cause
Other
Action
Landauer Issued an URGENT - Medical Device Recall notice to its consignees on 11/15/2023 via email. The notice explained the issue, potential risk, and requested return of the device.
Quantity
584 units
Distribution
US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K192196"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3008426232
Firm city
Glenwood
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register