Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, INC: Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system, Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. X-ray generator POLYDOROS A100 Plus

Recall number
Z-0614-2018
Recalling firm
Siemens Medical Solutions USA, INC
Product
Artis oneFloor-mounted system for uncompromised imaging; Artis one, Interventional fluoroscopic x-ray system, Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. X-ray generator POLYDOROS A100 Plus
Status
Terminated
Initiated
2017-08-22
Posted
not on file
Terminated
2018-08-21
Reason
The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screen
Root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent- Medical Device Correction letter dated August 22, 2017. Our service organization will contact you shortly to arrange a date to perform this corrective action (A system hardware update will be implemented and all affected DP Transceivers below revision 03 will be updated). Following the system hardware update, the cause is eliminated and a recurrence of the fault is prevented. For further questions, please call (610) 448-6461.
Quantity
207 worldwide (13 distributed in US)
Distribution
US Distribution to the states of : AR, MO, NY, IN, AR, TX , NC, PA, MI, LA, WA, CA and AL.
Lot, serial or model codes
Serial # 82044; 82105; 82070; 82054; 82040;82038;82065 82108; 82010; 82039; 82109 82102 82066
510(k) numbers
["K133580"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register