Roche Molecular Systems, Inc.: COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
- Recall number
- Z-0614-2013
- Recalling firm
- Roche Molecular Systems, Inc.
- Product
- COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
- Status
- Terminated
- Initiated
- 2012-10-11
- Posted
- 2012-12-23
- Terminated
- 2013-09-11
- Reason
- Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.
- Root cause
- Device Design
- Action
- Roche sent an Urgent Medical Device Correction notification/Fax Back forms dated 10/11/12 to potentially affected customers. The letter identified the affected product, the issue, clinical significance, workaround, and required actions, along with contact information. Customers are to use the workaround until their Roche FSE has performed a Master Initialization on their affected instruments. Also, the letter should be filed for future reference. Questions should be directed to Roche Molecular Diagnostics Technical Support at 1-800-526-1247.
- Quantity
- 7 units
- Distribution
- Distributed in the states of GA, CA, NC, and MN.
- Lot, serial or model codes
- CAP 394436 to 394663
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JJH
- Firm FEI
- 2243471
- Firm city
- S Branchburg
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28