Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Percept PC Implantable Neurostimulator (INS) - Product Usage: The Percept PC is a neurostimulation device for deep brain stimulation therapy. It is indicated for DBS treatment for Parkinson s disease, essential tremor, dystonia, obsessive-compulsive disorder, and epilepsy.

Recall number
Z-0612-2021
Recalling firm
Medtronic Neuromodulation
Product
Percept PC Implantable Neurostimulator (INS) - Product Usage: The Percept PC is a neurostimulation device for deep brain stimulation therapy. It is indicated for DBS treatment for Parkinson s disease, essential tremor, dystonia, obsessive-compulsive disorder, and epilepsy.
Status
Terminated
Initiated
2020-11-19
Posted
not on file
Terminated
2023-04-03
Reason
A software anomaly in the A620 Patient Programmer application was identified that results in failure to connect with the Percept PC device.
Root cause
Software design
Action
Beginning November 19, 2020 accounts impacted by this issue were verbally notified to quarantine affected product. Beginning November 25, 2020 a recall notification letter and retrieval instructions titled "URGENT: MEDICAL DEVICE RECALL A620 Patient Programmer issue with specific Percept" PC Neurostimulators" was hand delivered to impacted accounts by Medtronic Representatives.
Quantity
12 devices
Distribution
US Nationwide distribution including in the states of Arizona, California, New Hampshire, Pennsylvania, Tennessee, Texas, Washington.
Lot, serial or model codes
Model Number: B35200; GTIN: 00763000253363; Serial Numbers: NPI706941H, NPI707024H, NPI707000H, NPI706992H, NPI706963H, NPI706962H, NPI706987H, NPI706945H, NPI706952H, NPI706954H, NPI706955H, NPI706957H
510(k) numbers
not on file
PMA numbers
not on file
Product code
MHY
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register