FDA Medical Device Recalls (openFDA)
Enztec Limited: REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
- Recall number
- Z-0612-2020
- Recalling firm
- Enztec Limited
- Product
- REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
- Status
- Terminated
- Initiated
- 2019-08-16
- Posted
- 2020-01-24
- Terminated
- 2023-03-02
- Reason
- The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.
- Root cause
- Unknown/Undetermined by firm
- Action
- On August 16, 2019, a Urgent Medical device recall for Product Correction" letter was emailed to affected consignees. In addition to informing the consignees about the recall, the
- Quantity
- 312 Femoral Impactors
- Distribution
- CA
- Lot, serial or model codes
- Lot number Distribution Date: 161471 08-Jun-16, 161472 08-Jun-16, 161473 08-Jun-16, 161474 08-Jun-16, 161475 08-Jun-16, 161476 08-Jun-16, 161477 08-Jun-16, 161756 29-Jun-16, 161757 29-Jun-16, 161758 29-Jun-16, 161759 29-Jun-16, 161760 29-Jun-16, 161761 29-Jun-16, 161762 29-Jun-16, 161822 22-Jul-16, 161823 22-Jul-16, 161824 22-Jul-16, 161825 22-Jul-16, 161826 22-Jul-16, 161827 22-Jul-16, 161828 22-Jul-16, 170760 24-Mar-17, 170761 24-Mar-17, 170762 24-Mar-17, 170763 24-Mar-17, 170764 24-Mar-17, 170765 24-Mar-17, 170766 24-Mar-17, 170767 24-Mar-17, 170768 24-Mar-17, 171989 23-Jun-17 ,
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWA
- Firm FEI
- 3004641308
- Firm city
- Christchurch
- Firm state
- Sockburn
- Firm country
- New Zealand
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28