Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Rad Laboratories Inc: Bio-Rad Laboratories Autoimmune EIA ANA Screening kit, 96 test, model 96AN, Distributed and Manufactured by Bio-Rad Laboratories, Inc.

Recall number
Z-0612-2011
Recalling firm
Bio-Rad Laboratories Inc
Product
Bio-Rad Laboratories Autoimmune EIA ANA Screening kit, 96 test, model 96AN, Distributed and Manufactured by Bio-Rad Laboratories, Inc.
Status
Terminated
Initiated
2010-06-24
Posted
2010-12-13
Terminated
2011-08-08
Reason
Product mislabeled-- the ANA cutoff Control and Prostiva controls were mislabeled. Use of the incorrect controls will cause run failure - patient results cannot be reported because controls will not pass..
Root cause
Labeling mix-ups
Action
Bio-Rad issued Medical Device Recall Notification letters dated June 24, 2010 to consignees, identifying the affected products and requested that all affected product be disposed of in accordance with local procedures. Response forms are to be completed and returned to the firm. Bio-Rad can be contacted for replacement product and about this recall at 510 741-6114.
Quantity
37 units
Distribution
Nationwide Distribution: including the states of California, Ohio, Massachusetts, and Missouri. Product delivered to 11 consignees only in the US.
Lot, serial or model codes
Manufactured on 06/07/2010, lot number 24501097
510(k) numbers
["K954723"]
PMA numbers
not on file
Product code
LKJ
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register