FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.

Recall number
Z-0612-06
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.
Status
Terminated
Initiated
2006-02-07
Posted
2006-03-16
Terminated
2008-03-24
Reason
A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply form confirming receipt of the letter. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Quantity
282 pumps
Distribution
Nationwide. Arizona, California, Florida, Georgia, Hawaii, Iowa, Indiana, Kentucky, Louisiana, Michigan, Mississippi, North Carolina, Nebraska, New Mexico, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, Washington, West Virginia
Lot, serial or model codes
serial numbers 15060001AB, 15060002AB, 15060004AB/15060013AB, 15060015AB/15060021AB, 15060023AB/15060073AB, 15060075AB, 15060077AB/15060116AB, 15060116AB, 15060118AB/15060126AB, 15060128AB/15060130AB, 15060132AB/15060172AB, 15060174AB/15060209AB, 15070003AB/15070006AB, 15070008AB/15070018AB, 15070021AB, 15070023AB, 15070026AB/15070034AB, 15070037AB15070041AB, 15070044AB/15070047AB, 15070049AB/15070052AB, 15070056AB/15070059AB, 15070066AB, 15070069AB, 15070071AB, 15070072AB, 15070076AB, 15070077AB, 15070079AB, 15070081AB, 15070082AB, 15070084AB, 15070086AB, 15070087AB, 15070092AB, 15070094AB/15070096AB, 15070099AB, 15070101AB, 15070109AB, 15070111AB/15070115AB, 15070117AB, 15070119AB, 15070121AB, 15070122AB, 15070124AB, 15070126AB/15070128AB, 15070130AB/15070132AB, 15070138AB/15070140AB, 15070144AB
510(k) numbers
["K945942"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register