KCI USA, Inc.: RotoProne Therapy System, Model 209500, Device List Number E232632 Product is designed to allow patients to be mechanically rotated to the prone position and provides kinetic therapy to patients. It is intended for the treatment and prevention of pulmonary complications associated with immobility.

Recall number
Z-0611-2011
Recalling firm
KCI USA, Inc.
Product
RotoProne Therapy System, Model 209500, Device List Number E232632 Product is designed to allow patients to be mechanically rotated to the prone position and provides kinetic therapy to patients. It is intended for the treatment and prevention of pulmonary complications associated with immobility.
Status
Terminated
Initiated
2010-09-09
Posted
2010-12-13
Terminated
2011-03-21
Reason
Product complaints indicating that while attempting to resolve an alarm or error condition, operator was unable to return product to the prone position for patient.
Root cause
Other
Action
KCI USA, Inc. sent an URGENT - Medical Device Safety Alert letter, dated September 8, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the cusotmer. Customers were instructed to have a practiced back-up plan, such as a plan for the manual proning of patients in the event that clinical circumstances dictate the need for more immediate prone therapy. For questions, customers in the US should call KCI's Technical Support Center at 1-800-275-4524, option 5. For customers outside the US they should contact their local sales representative. Customers should contact their accounts with affected product and advise them of the safety notice.
Quantity
240 units
Distribution
Worldwide Distribution - USA Nationwide, United Kingdom, Germany, and Saudi Arabia
Lot, serial or model codes
Serial number ranges: RRFK00036-RRFK00047; RRFK00049-RRFK00115; RRFK00117-RRFK00128; RRFK00141-RRFK00211; RRFK00213-RRFK00233; RRFK00235; RRFK00237-RRFK00254; RRFK00256-RRFK00270; RRFK00273-RRFK00280; RRFK00282-RRFK00284; RRFK00287-RRFK00290; RRJK0001; RRIK00003; RRIK00004; RRFK00234; RRFK00236; RRFK00285; RRFK00286; and RRJK00002
510(k) numbers
not on file
PMA numbers
not on file
Product code
IKZ
Firm FEI
1625774
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register