Skip to the content

FDA Medical Device Recalls (openFDA)

OMNI LIFE SCIENCE: OMNIBotics Tracker Kit -Model: RM-10000 The TOTAL KNEE SURGETICS Navigation System with iBlock (Now OMNIBotics) is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. It is specifically indicated for: Total Knee Arthroplasty

Recall number
Z-0610-2017
Recalling firm
OMNI LIFE SCIENCE
Product
OMNIBotics Tracker Kit -Model: RM-10000 The TOTAL KNEE SURGETICS Navigation System with iBlock (Now OMNIBotics) is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. It is specifically indicated for: Total Knee Arthroplasty
Status
Terminated
Initiated
2016-10-14
Posted
2016-11-17
Terminated
2017-02-01
Reason
Incorrect lot number on outer kit
Root cause
Error in labeling
Action
The firm, OMNIlife Science, Inc., issued an "URGENT MEDICAL DEVICE ADVISORY NOTICE" dated 10/14/16 by email to its consignees/customers. U.S. and foreign agents were also sent the notification packet via overnight mail. Acknowledgement of receipt is requested. The notice described the product, problem and actions to be taken. The customers were instructed to immediately quarantine lot 19-09-021 for OMNIBotics Tracker Kit RM-1000; overlabel the outside packaging with the label provided following the provided training paperwork and rework instructions; note advisory notice only includes listed product, it does not affect other lots; and complete and return the Acknowledgement and Receipt Form and accompanying FN-102 by Fax to 508-519-3390. Accounts are to hold item RM-10000 with the lot # 19-09-021 until a corrected label is provided for the outer box. The patient labels inside the box are correct. For further information , Regulatory Affairs Specialist at 774-226-1847 or 800-448-6664 Monday through Friday, 8:00 AM to 4:30 PM, Easter Time.
Quantity
15 units
Distribution
Worldwide Distribution: US distribution to OK, MD, TX and country of: Australia.
Lot, serial or model codes
Lot number: 19-09-021
510(k) numbers
["K090953"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3008342029
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register