Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Versys Build-Up Block The VerSys CRC Hip System is indicated for total hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by defi ciencies of the femoral head, neck, or portions of the proximal femur. Arthroplasty should be performed only when more conservative methods of treatment have failed to provide symptomatic relief and when there is progressive disability.

Recall number
Z-0610-2014
Recalling firm
Zimmer, Inc.
Product
Versys Build-Up Block The VerSys CRC Hip System is indicated for total hip arthroplasty in patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by defi ciencies of the femoral head, neck, or portions of the proximal femur. Arthroplasty should be performed only when more conservative methods of treatment have failed to provide symptomatic relief and when there is progressive disability.
Status
Terminated
Initiated
2013-12-10
Posted
2013-12-30
Terminated
2014-09-11
Reason
Zimmer is initiating a lot specific recall of the CPT Hip Stem, Versys Build-Up Block, and Trabecular Metal Reverse Glenosphere medical devices due to reports of the products being packaged in the incorrect outer carton. Product labels and patient labels are correct.
Root cause
Labeling mix-ups
Action
Zimmer sent an Urgent Medical Device Recall letter dated December 10, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification and ensure affected personnel were aware of the contents. If affected product was found, it should be quarantines and their Zimmer representative notified. The Zimmer sales representative will remove the recalled product from their facility. Customers with questions were instructed to call 1-877-946-2761. .
Quantity
3
Distribution
Worldwide Distribution -USA including OH, FL, WI, MI NJ, SC, CA, NC, PA, OR, WA, TN, KY, OK, KS, AZ, TX, CO, NV and Internationally to Germany. There have been 71 distribution events. Note that this product is lot controlled and may be distributed multiple times if it was returned and returned to the shelf. 7 units have been returned, including three through the complaint process, which leaves 64 affected units in the field subject to removal.
Lot, serial or model codes
Part 00-7871-001-10 Lot 62359872
510(k) numbers
["K913649","K960658"]
PMA numbers
not on file
Product code
JDI
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register