Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

Recall number
Z-0609-2011
Recalling firm
Zimmer Inc.
Product
NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.
Status
Terminated
Initiated
2010-04-16
Posted
2010-12-13
Terminated
2012-09-12
Reason
The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery. These complications may also require revision surgery.
Root cause
Device Design
Action
Zimmer issued "Urgent Medical Device Safety Alert" letters dated April 16, 2010 to Risk Managers, user facilities, Health Care Professionals and Zimmer Sales force. The users were informed to inspect the device for loosening. The alert also informed users to inspect the device after each use to verify that the device was intact. Customers were instructed to complete and return a User Facility Documentation Form. Zimmer will repair and replace all affected devices. Customers can contact ZImmer, Inc. at 800 613-6131 concerning this recall.
Quantity
10,666 all products.
Distribution
Worldwide Distribution: Throughout US including Puerto Rico, and the countries of Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Denmark, Ecuador, El Salvador, France, Germany, Guatemala, Hong Kong, India, Italy, Japan, Korea, Mexico, Netherlands, Nicaragua, Panama, Peru, Portugal, Singapore, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine, and United Kingdom.
Lot, serial or model codes
61650800, 64173900, 64314000, 64592700, 64899900, 65307200, 66384700, 67271500, 68001000, 68507500, 70066700, 72635200, 73040600, 74330100, 78288600, 78775900, 79027000, 79301900, 79414800, 79911900, 50006600, 50243100, 50435900, 51091100, 05478400, 52676400, 53616000, 53780600, 54122000, 54289900, 54601600, 55023000, 55023000, 55307800, 56221200, 56599400, 56760200, 56795200, 56934500, 56974400, 57038500, 57111600, 57128100, 57241400, 57440800, 57329200, 57510900, 57543000, 57575000, 57645600, 57854100, 57918600, 57808100, 58005700, 58318100, 58318200, 58575700, 58812400, 58784400, 59456600, 59183600, 59671000, 59660300, 59786900, 60481100, 60844300, 60903700, 60939100, 61813100, 62260000, 05631100, 62994200, 64835300, 65403600, 65861500, 66614100, 66883200, 67441900, 67599300, 68466000, 68764200, 69175100, 69286900, 70153900, 70294600, 70658600, 71027500, 71280200, 71572500, 71866600, 72502300, 72829600, 73433500, 74078000, 74644700, 74750900, 75331100, 75504400, 76422600, 79626500, 41996400, 52288600, 53220100, 54000500, 60007281, 60025415, 60029968, 60034339, 60056415, 60066606, 60064664, 60077494, 60079251, 60091628, 60092158, 60102687, 60105706, 60115562, 60117109, 60120485,  60124571, 60137142, 60149310, 60160018, 60174138, 60194573, 60195084, 60209845, 60219302, 60224592, 60226046, 60225877, 60240482, 60271491, 60278677, 60280712, 60302621, 60314459, 60328822, 60341569, 60359101, 60362634, 60366062, 60378347,  60380488, 60398085, 60401003, 60425875, 60439542, 60451190, 60523326, 60547208, 60565239, 60568430, 60582612, 60587278, 60603208, 60652168, 60685947, 60723702, and 60781789.
510(k) numbers
["K042271","K072619","K933785","K960279","K991581"]
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register