Skip to the content

FDA Medical Device Recalls (openFDA)

Sanofi Genzyme: Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe Product Usage: A elastoviscous high molecular weight fluid containing hylan A and hylan B polymers produced from chicken combs. Synvisc-One is administered as a single intra-articular injection. Synvisc-One is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics, e.g., acetaminophen.

Recall number
Z-0607-2018
Recalling firm
Sanofi Genzyme
Product
Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe Product Usage: A elastoviscous high molecular weight fluid containing hylan A and hylan B polymers produced from chicken combs. Synvisc-One is administered as a single intra-articular injection. Synvisc-One is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics, e.g., acetaminophen.
Status
Terminated
Initiated
2017-12-11
Posted
not on file
Terminated
2023-04-03
Reason
Contamination with Methylbacterium thiocyanatum
Root cause
Environmental control
Action
On December 11, 2017 Sanofi Genzyme distributed URGENT VOLUNTARY DEVICE RECALL notices and reply cards to their customers via courier service. Customers were advised to inspect inventory, immediately discontinue use and return all affected product. On December 19, 2017 Sanofi Genzyme distributed a 2nd notice titled DEVICE RECALL UPDATE NOTICE (2 of 3) notifying customers that the isolated bacterial organism was Methylbacterium thiocyanatum. On December 22, 2017 Sanofi Genzyme distributed a 3rd notice titled DEVICE RECALL UPDATE NOTICE (3 of 3) which included antibiotic sensitivity analysis guidance for the isolated organism. Customers who wish to return product should contact GENCO Pharmaceutical Services at 1-855-838-5782 for a recalled product return kit. Customers are encouraged to complete and return the Business Reply Card via fax to 1-414-459-8791 or via email to recalls@genco.com. Sanofi Genzyme will issue you a credit or replacement product upon receipt of the recalled product by GENCO. All questions of a medical or clinical nature should be directed to Sanofi Medical Information Services in the U.S. at 1-800-633-1610, option 1. Sanofi US Customer service can be reached at 1-800-633-1610 option 7 then 4.
Quantity
12,380 units
Distribution
US Nationwide distribution(36 states)
Lot, serial or model codes
Lot Number: 7RSL021 Expiration Date: 2020-05-31
510(k) numbers
not on file
PMA numbers
["P940015"]
Product code
MOZ
Firm FEI
1000305672
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register