FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: SOMATOM Definition Edge, Computed tomography x-ray system The Siemens SOMATOM Definition AS/ AS+ (Project P46) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles
- Recall number
- Z-0607-2017
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- SOMATOM Definition Edge, Computed tomography x-ray system The Siemens SOMATOM Definition AS/ AS+ (Project P46) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles
- Status
- Terminated
- Initiated
- 2016-11-02
- Posted
- not on file
- Terminated
- 2017-11-08
- Reason
- Siemens is releasing a software update that provides bug-fixes to improve system performance for customers with SOMATOM Definition AS, SOMATOM Definition Edge, SOMATOM Definition Flash Systems with software version VA48A_SP2 and Care Contrast license correction of potential safety issues.
- Root cause
- Software design
- Action
- Siemens mailed a Customer Safety Notice on November 9, 2016, to customers who were affected by the recall to inform them of a possible malfunction with the systems. The notification explained the underlying issue, the effect if has on the operation of the system, potential risks involved; and the actions they plan to implement to correct the issues. Customers with questions were instructed to call 1-800-888-7436. For questions regarding this recall call 610-448-6471.
- Quantity
- 7 systems
- Distribution
- Nationwide distribution to MI, NY, CA, KY, ND, and NE.
- Lot, serial or model codes
- Model Number: 1059000 Serial Numbers:96020, 96047,74192,83398,96011,83387 96029
- 510(k) numbers
- ["K152036"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as