Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems, Inc.: Segmented Cervix Applicator Set, part number GM11004310, a component used with the tube for segments and cervical sleeves The guide tube for segments and cervical sleeves in the Segmented Cervix Applicator Set was developed to treat cancer of the cervix, vagina, endometrium, vaginal stump and uterus. It is used for Brachytherapy.

Recall number
Z-0607-2013
Recalling firm
Varian Medical Systems, Inc.
Product
Segmented Cervix Applicator Set, part number GM11004310, a component used with the tube for segments and cervical sleeves The guide tube for segments and cervical sleeves in the Segmented Cervix Applicator Set was developed to treat cancer of the cervix, vagina, endometrium, vaginal stump and uterus. It is used for Brachytherapy.
Status
Terminated
Initiated
2012-09-28
Posted
2012-12-20
Terminated
2013-04-02
Reason
The guide tube which is part of the Segmented Cervix Applicator Set is difficult to disconnect from cervical sleeve.
Root cause
Device Design
Action
The firm, Varian Medical Systems, sent an "URGENT MEDICAL DEVICE CORRECTION - URGENT FIELD SAFETY NOTICE" letter dated September 28, 2012 to all affected consignees/Customers. The letter identified the affected product, description of problem and recommended user actions to be taken. The consignees/customers were instructed to immediately remove all affected guiding tubes from clinical use, and sequester the components for pending return to Varian Brachy Therapy. Consignees/Customers were requested to complete the attached proof of notification form and return it to Varian Medical Systems. For questions contact-AT&T international direct dialing network (http://www.business.att.com/bt/dial_guide.isp. Then after hearing the AT&T signature noise, and. "Please enter the number you are calling", dial 800-360-7909; UK:+44 1293-601-327 or US:+1-434-979-1540. To submit a support request or general inquiry, compose an email with a blank subject line to: brachyhelp@varian.com.
Quantity
15 units
Distribution
Worldwide Distribution - U.S. state of Georgia and the countries of Canada, Germany, India, Poland, and Thailand.
Lot, serial or model codes
Lot L33, Serial numbers H64Y009-GM, H64A056, H640576, H600272, H640784, H640774, H640764, H600603, H640547.
510(k) numbers
["K983436"]
PMA numbers
not on file
Product code
JAQ
Firm FEI
3000206172
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register