FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: Access Immunoassay Systems, Part Number: 81600 ; Access 2 Immunoassay Systems, Part Number: 81600N; Synchron LXi 725; UniCel DxI 800 Access Immunoassay Systems, Part Number: 973100; UniCel DxI 600 Access Immunoassay Systems, Part Number A30260; UniCel DxC 600i, Part Number: A25656; UniCel DxC 660i, Part Number: A64871; UniCel DxC 860i, Part Number: A64935; UniCel DxC 880i, Part Number: A59102; SYNCHRON Access Clinical Systems The Access/DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentration found in human body fluids. Function of the Peristaltic Pump: to aspirate unbound material from reaction vessels and to move waste from vacuum jar into waste bottle through associated tubing within the Access Immunoassay System.
- Recall number
- Z-0607-2011
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access Immunoassay Systems, Part Number: 81600 ; Access 2 Immunoassay Systems, Part Number: 81600N; Synchron LXi 725; UniCel DxI 800 Access Immunoassay Systems, Part Number: 973100; UniCel DxI 600 Access Immunoassay Systems, Part Number A30260; UniCel DxC 600i, Part Number: A25656; UniCel DxC 660i, Part Number: A64871; UniCel DxC 860i, Part Number: A64935; UniCel DxC 880i, Part Number: A59102; SYNCHRON Access Clinical Systems The Access/DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentration found in human body fluids. Function of the Peristaltic Pump: to aspirate unbound material from reaction vessels and to move waste from vacuum jar into waste bottle through associated tubing within the Access Immunoassay System.
- Status
- Terminated
- Initiated
- 2010-08-13
- Posted
- 2010-12-13
- Terminated
- 2012-07-11
- Reason
- This recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated as a result of missing or incorrect assay reagents. In these cases both false positive and false negative results are possible, depending upon the assay configuration.
- Root cause
- Pending
- Action
- Beckman Coulter, Inc. sent two (2) Product Corrective Action (PCA) letters (Access letter and DxI letter) with attached Customer Response form the week of August 09, 2010 to the affected customers. The letters were dated August 13, 2010, and identified the product, the problem, and the action to be taken by the customers. The letter reminded customers that it is critical that they load reagent packs according to the procedure provided in their Access or Access 2 Operators Guide, Instructions for Use, or Help system. A copy of the procedure was included with the letter. Customers were reminded that all operators should be properly trained on these important procedures and understand the impact on reported results if these procedures are not followed. Customers were instructed to share this information with their laboratory staff, and retain this notification as part of their laboratory Quality System documentation. Customers who needed assistance or had any questions regarding this notification, were instructed to contact Technical Support at 1-800-854-3633, option 1 in the United States or Canada.
- Quantity
- 9571 units (instruments)
- Distribution
- Nationwide Distribution & Canada
- Lot, serial or model codes
- All versions are affected.
- 510(k) numbers
- ["K023764","K060256","K922823"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28