Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson and Company: Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx For Intraocular Use REF 585304 Use as a surgical aid to protect corneal endothelium during cataract extraction procedures, intraocular lens implantation and anterior segment surgery.

Recall number
Z-0606-2011
Recalling firm
Becton Dickinson and Company
Product
Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx For Intraocular Use REF 585304 Use as a surgical aid to protect corneal endothelium during cataract extraction procedures, intraocular lens implantation and anterior segment surgery.
Status
Terminated
Initiated
2010-10-04
Posted
2010-12-16
Terminated
2012-02-16
Reason
Finger flange on the BD 1% OVD syringe can disengage during cataract surgery and and may lead to injury.
Root cause
Component design/selection
Action
Beaver Visitec issued an Urgent: Field Safety Notice letter, dated 10/4/10 to direct accounts, identifying the affected product and informing users of the potential for the finger grip to disengage during use. The firm offers solutions as follows: Allow the finger grip to remain attached to the syringe and continue to use the product. Position the open side of the finger grip towards the palm to avoid detachment of the grip from the syringe. Remove the finger grip and continue to use the product. Return the unused product to Beaver-Visitec for reimbursement. Users were requested to complete the Customer Response Form. Questions can be directed to: Inside Sales at 781-906-7952.
Quantity
72278 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot No. Exp. Date RDOl71B 2012-12; RD0070A 2012-05; RD0018A 2012-02; RD0007A 2012-01 RC0064C 2011-06; RC0004A 2011-01;
510(k) numbers
not on file
PMA numbers
["P960011"]
Product code
LZP
Firm FEI
1211998
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register