Abbott Medical Optics Inc (AMO): AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)

Recall number
Z-0605-2010
Recalling firm
Abbott Medical Optics Inc (AMO)
Product
AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)
Status
Terminated
Initiated
2009-11-11
Posted
2010-01-12
Terminated
2012-07-25
Reason
Abbott Medical Optics (AMO) has received complaints of some Tecnis 1-Piece IOLs (model ZCB00) have adhered to the lens optic after insertion into the eye. This can lead to a need for additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.
Root cause
Process design
Action
Beginning on Friday, November 6, 2009, AMO began distribution of the Recall Notification Letter via Federal Express to the 874 AMO Tecnis 1-piece IOL customer accounts in the US. Included with the Recall Notification letter is a facsimile delivery confirmation that each customer was instructed to fax back to AMO to confirm receipt of the Recall Notification Letter. --- Beginning on Monday, November 9, 2009, AMO began distribution of the Recall Notification letter to the AMO Tecnis 1-piece IOL customer/distributors accounts outside the USA. "THIS RECALL NOTIFICATION DOES NOT INVOLVE TECNIS 1-PIECE IOL THAT HAVE BEEN IMPLANTED. ONCE THE HAPTICS HAVE BEEN SUCCESSFULLY DEPLOYED, THESE LENSES FUNCTION PROPERLY." Questions should be directed to an AMO Customer Service Representative at 1-877-AMO-4LIFE (1-877-266-4543).
Quantity
246000 Lenses
Distribution
AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, Hawaii, and Puerto Rico. INTERNATIONAL: Austria, Azerbaijan, Belgium, Canada, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Italy, Lebanon, Lichtenstein, Libya, Martinique, Netherlands, Poland, Palestine, Portugal, Saudi Arabic, Sweden, Slovenia, Turkey, and South Africa.
Lot, serial or model codes
All Serial numbers of tecnis 1-Piece Intraocular Lenses that were processed through lens surface treatment at the AMO PR facility between May 19 and October 16, 2009 - with serial numbers ending in 0906, 0907, 0908, and 0909. and subset with serial numbers ending in 0905 and 0910.  Expiration/Use By Date (YYYY-MM): 2011-05, 2011-06, 2011-07, 2011-08, 2011-09, and 2011-10
510(k) numbers
not on file
PMA numbers
["P980040S015"]
Product code
HQL
Firm FEI
3003843509
Firm city
Santa Ana
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register