Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Corporation: NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK;

Recall number
Z-0601-2026
Recalling firm
Stryker Corporation
Product
NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK;
Status
Open, Classified
Initiated
2025-10-31
Posted
2025-11-26
Terminated
not on file
Reason
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Root cause
Under Investigation by firm
Action
On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.
Quantity
39,148
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
Lot, serial or model codes
Product Number/UDI-DI (GTIN)/Lot Numbers: NN-8024 00816744028244 37AI2944 37AJ2013 37BD3753 37BD3778 37CA1605 37CB3738 37CH0636 37CI1919 37EH0405 37FC3059 37FD0154 37GA4512 37GH1715 37GI0110 37GJ1623 37HB0029 37HH2313 37HI1536 37HI1631 37IA7694 37ID4162 37IG2900 37JB1413 37JH0131 37JH2903 37KG1406 37KH2953 37LA0708 37LG1834; NN-8025 00816744028251 37BB2326 37BJ1836 37CH0640 37DC0545 37HD2766 37KB3677 37KJ1139; NN-8026 00816744028268 37BJ1837 37CC7852 37CH0643 37DJ1056; NN-8042 07613327663280 37CC9220 37ED9408 37FJ2202 37II2308 37LD8666
510(k) numbers
["K191599"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register