Skip to the content

FDA Medical Device Recalls (openFDA)

Hill-Rom Manufacturing, Inc.: Synergy Dynamic, Powered Patient Rotation Bed, Model number 5607. Designed to provide enhanced static, pulsation and rotational pressure relieving therapies. Provides sixteen pressure zones for customizing the support surface to each patient, and is capable of providing alternating pressure therapy and lateral rotation up to 45 degrees side to side.

Recall number
Z-0601-2009
Recalling firm
Hill-Rom Manufacturing, Inc.
Product
Synergy Dynamic, Powered Patient Rotation Bed, Model number 5607. Designed to provide enhanced static, pulsation and rotational pressure relieving therapies. Provides sixteen pressure zones for customizing the support surface to each patient, and is capable of providing alternating pressure therapy and lateral rotation up to 45 degrees side to side.
Status
Terminated
Initiated
2008-10-09
Posted
2009-01-07
Terminated
2009-12-10
Reason
These devices have the potential to overheat.
Root cause
Device Design
Action
All end users (those who either purchased the device and any customer that has rented) were notified by an Urgent Field Safety Notice on/about 10/31/2008. The letter informed the customers of the safety risk and list steps they can take to mitigate the risk until a Hill-Rom representative is able to modify the device. For the devices that were sold, the consignee of each device was notified by letter for the safety risk and what they can do to mitigate the risk until a Hill-Rom representative is able to modify the device. For the devices that are rented, an internal notification was sent via e-mail to Service Centers (Because the device is owned by the manufacturer, Hill-Rom, each Service Center in possession of the device will be the Consignee). Each Service Center will perform the correction on the units affected.
Quantity
2,083 units (Total all four units)
Distribution
Nationwide Distribution.
Lot, serial or model codes
All serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNM
Firm FEI
1045510
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register