Cook Inc.: Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL0-HC, Order Number G56221 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.

Recall number
Z-0600-2021
Recalling firm
Cook Inc.
Product
Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL0-HC, Order Number G56221 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.
Status
Terminated
Initiated
2020-11-24
Posted
2020-12-23
Terminated
2022-09-07
Reason
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
Root cause
Process control
Action
The firm, Cook Medical, sent "URGENT: MEDICAL DEVICE RECALL" letters dated 11/24/20 to customers. The letter described the product, problem and actions to be taken. The customers were instructed to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This medical device recall letter must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing devices potentially impacted from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue. Other Information This action is being taken with the knowledge of the Food and Drug Administration. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to
Quantity
4408
Distribution
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, KR, Latvia, Lithuania, MY, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, TW, TH, and United Arabs Emirates.
Lot, serial or model codes
Lot Numbers: 13143098, 13143099, 13143100, 13145952, 13145953, 13145955, 13150444, 13150446, 13159696, 13159697, 13159706, 13159708, 13166164, 13166165, 13166168, 13175473, 13175474, 13175480, 13178582, 13186336, 13192068, 13192074, 13211826, 13211827, 13214440, 13217125, 13217126, 13221802, 13223830, 13223831, 13229680, 13231823, 13231824, 13240026, 13240027, 13243203, 13246547, 13246566, 13252353, 13260022, 13260023, 13265330, 13265331, 13265335, 13268038, 13279080, 13281540, 13281800, 13281801, 13286656, 13286658, 13286659, 13286660, 13286662, 13288115, 13304447, 13306333, 13306334, 13307178, 13307189, 13315200, 13315217, 13320585, 13320594, 13329232, 13329248, 13332167, 13334774, 13334775, 13342868, 13349026, 13352485, 13366188, 13366191, 13366194, 13366195, 13366196, 13369415, 13369426, 13380379, 13382712, 13391138, 13391139, 13396128, 13404831, 13404832, 13416497, 13420104, 13420105, 13281802X, NS13315219, NS13315220, NS13329249, NS13334773, NS13349037, NS13352484, NS13352486, NS13401845, NS13426336
510(k) numbers
["K142829"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register