FDA Medical Device Recalls (openFDA)
C.R. Bard, Inc.: Magic3 Antibacterial Hydrophilic Intermittent Catheter, Male 14 Fr. (30/Box), Sterile Product Usage: The catheter is an all silicone, single lumen, drainage tube with drainage eyes at the proximal end and a tapered funnel at the distal end. The outer surface of the catheter has an antibacterial and hydrophilic coating which binds water molecules to the surface creating a smooth lubricating film. It is packaged with a water sachet that is manually burst to hydrate the catheter coating at time of use. The catheters are packaged in a film-to-film pouch and are sterilized by electron beam irradiation and are distributed as a single use device. The catheter is intended for urinary bladder drainage in adult males and females requiring catheterization for management of incontinence, voiding dysfunction and surgical procedures. Efficacy of the catheter in preventing urinary tract infection during intermittent use has not been established. The device is not intended to be used as a treatment for active urinary tract infection. For urological use only; urinary catheters are intended for use for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.
- Recall number
- Z-0600-2017
- Recalling firm
- C.R. Bard, Inc.
- Product
- Magic3 Antibacterial Hydrophilic Intermittent Catheter, Male 14 Fr. (30/Box), Sterile Product Usage: The catheter is an all silicone, single lumen, drainage tube with drainage eyes at the proximal end and a tapered funnel at the distal end. The outer surface of the catheter has an antibacterial and hydrophilic coating which binds water molecules to the surface creating a smooth lubricating film. It is packaged with a water sachet that is manually burst to hydrate the catheter coating at time of use. The catheters are packaged in a film-to-film pouch and are sterilized by electron beam irradiation and are distributed as a single use device. The catheter is intended for urinary bladder drainage in adult males and females requiring catheterization for management of incontinence, voiding dysfunction and surgical procedures. Efficacy of the catheter in preventing urinary tract infection during intermittent use has not been established. The device is not intended to be used as a treatment for active urinary tract infection. For urological use only; urinary catheters are intended for use for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.
- Status
- Terminated
- Initiated
- 2016-10-05
- Posted
- not on file
- Terminated
- 2021-10-08
- Reason
- Misbranding; the product labeled as an Antibacterial Hydrophilic Intermittent Catheter, is instead an Antibacterial Intermittent Catheter (no Hydrophilic Coating).
- Root cause
- Mixed-up of materials/components
- Action
- Bard Medical sent an Urgent Medical Device Product Recall letter dated October 5, 2016 to customers The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine and quarantine and return affected inventory. Customers were asked to complete and return the accompanying Recall & Effectiveness Check Form attached.
- Quantity
- 10,290 units
- Distribution
- US Nationwide Distribution in the states of: CA, FL, IL, IN, MN, MO, NC, NJ, NY, OH, OK, PA, TX and UT.
- Lot, serial or model codes
- Product Code: 53514, Lot number: 53622813, Exp. Date: 30Sep2017.
- 510(k) numbers
- ["K033477"]
- PMA numbers
- not on file
- Product code
- MJC
- Firm FEI
- 1018233
- Firm city
- Covington
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28