Skip to the content

FDA Medical Device Recalls (openFDA)

C. R.Bard, Inc./Electrophysiology Division: Bard Stinger S Ablation Catheter: Item Numbers 210001S 218501S 6210015S 210002S 218502S 6210016S 210003S 218503S 6210017S 210004S 218504S 6210018S 210005S 218505S 6210019S 210006S 218506S 6210020S 210007S 218507S 210015S 218516S 210016S 218518S 210017S 218519S 210018S 218520S 210019S 218521S 210020S 218523S 210023S 218524S 210024S 218525S 210025S 218526S 210026S 218527S 210027S 218528S 210028S 218529S 210029S 210030S

Recall number
Z-0600-03
Recalling firm
C. R.Bard, Inc./Electrophysiology Division
Product
Bard Stinger S Ablation Catheter: Item Numbers 210001S 218501S 6210015S 210002S 218502S 6210016S 210003S 218503S 6210017S 210004S 218504S 6210018S 210005S 218505S 6210019S 210006S 218506S 6210020S 210007S 218507S 210015S 218516S 210016S 218518S 210017S 218519S 210018S 218520S 210019S 218521S 210020S 218523S 210023S 218524S 210024S 218525S 210025S 218526S 210026S 218527S 210027S 218528S 210028S 218529S 210029S 210030S
Status
Terminated
Initiated
2003-01-08
Posted
2003-02-28
Terminated
2012-05-03
Reason
Sterile barrier maybe perforated compromising sterility of the device
Root cause
Packaging process control
Action
Bard Electrophysiology notified direct accounts by letter on 1.08.03 via Fed''X. Accounts are requested to return product.
Quantity
not on file
Distribution
Nationwide
Lot, serial or model codes
All lot numbers were the 4th digit contains the letter A through M. Ex. 07A''A''1234
510(k) numbers
not on file
PMA numbers
not on file
Product code
not on file
Firm FEI
1000120754
Firm city
Billerica
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register