Siemens Healthcare Diagnostics, Inc.: Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
- Recall number
- Z-0599-2025
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
- Status
- Open, Classified
- Initiated
- 2024-10-21
- Posted
- 2024-12-04
- Terminated
- not on file
- Reason
- Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.
- Root cause
- Under Investigation by firm
- Action
- Siemens issued Urgent Medical Device Correction (UMDC) ACHC25-02.A.US was issued to US affected customers by FedEx beginning on 10/28/2024 and Urgent Field Safety Notice (UFSN) ACHC25-02.A.OUS. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Manually dilute patient samples that are > 18.00 mg/L: 1. Navigate to the CH Test Definition screen. 2. Under Measuring Interval, uncheck Repeat when Outside Measuring Interval. 3. For samples that generate test results > 18.00 mg/L, prepare a 1:2 manual dilution with saline solution (0.9% unbuffered Isotonic saline). 4. From the Create Patient Order screen, order the sample and enter 2 in the manual dilution field on the left of the screen. Note: The system will apply the 1:2 dilution factor and calculate the final result. 5. Process the manually diluted sample. "Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. "Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution The Atellica CH B2M assay IFU will be updated with the above steps to support manual 1:2 dilution only with saline for samples > 18.00 mg/L. Please see Appendix, Table 2, for data using these revised instructions. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Quantity
- 11330 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.
- Lot, serial or model codes
- UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JZG
- Firm FEI
- 2432235
- Firm city
- Tarrytown
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28