Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) and product codes 1M8560 and 1M8560R (pumps without drug library option); Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.

Recall number
Z-0599-06
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) and product codes 1M8560 and 1M8560R (pumps without drug library option); Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.
Status
Terminated
Initiated
2006-02-08
Posted
2006-03-07
Terminated
2010-01-19
Reason
The occlusion detection and alarm may be delayed when the pumps are used with larger volume syringes at low flow rates.
Root cause
Other
Action
Important Product Information letters dated 2/8/06 were sent to the AS40 customers via first class mail to the attention of the Director of Nursing. The accounts were informed of the delay in the detection of a downstream occlusion when the AS40 pump is used with large volume syringes at low flow rates. To minimize the occlusion detection times at low flow rates, Baxter recommended that the customers take the following actions: * Use the smallest possible syringe size for the volume of fluid to be delivered; * Select the low occlusion pressure setting to minimize the detection time; * Utilize infusion sets with thick-walled, small bore tubing to minimize the detection time; and * Always verify that the clamping device or stopcock is open in the fluid path prior to starting an infusion.
Quantity
17,400 pumps
Distribution
Nationwide and Canada
Lot, serial or model codes
all serial numbers
510(k) numbers
["K940147"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register