FDA Medical Device Recalls (openFDA)
Cartiva, Inc: Cartiva Synthetic Cartilage Implant (SCI) Catalog: CAR-06-US (6mm), CAR-08-US (8mm), CAR-10-US (10mm), CAR-12-US (12mm)
- Recall number
- Z-0598-2025
- Recalling firm
- Cartiva, Inc
- Product
- Cartiva Synthetic Cartilage Implant (SCI) Catalog: CAR-06-US (6mm), CAR-08-US (8mm), CAR-10-US (10mm), CAR-12-US (12mm)
- Status
- Open, Classified
- Initiated
- 2024-10-31
- Posted
- 2024-12-04
- Terminated
- not on file
- Reason
- Patients implanted with synthetic cartilage implant, may experience a higher-than expected occurrence rate of the following hazards: revision, removal, implant subsidence, displacement, pain, nerve damage or fragmentation.
- Root cause
- Device Design
- Action
- On 10/31/24 recall notices were mailed or emailed to customers asking them to do the following: 1) Continue to follow patients treated with an impacted product for new or worsening symptoms of pain, difficulty walking, skin reactions, stiffness, swelling, or weakness of the big toe joint, consistent with your follow up protocols. Per Instructions for Use: the long-term effects of cartilage replacement are not known; and the clinical and medical status of each patient should be considered when treating patients. 2) To help minimize complications, reference the information in the Instructions for Use and the information included in this notification. Per standard practice, continue to discuss all potential risks identified and discuss the benefits and risks of all relevant treatment options for first metatarsophalangeal joint osteoarthritis with your patients. 3) Check your internal inventory to locate the products listed on the attached business reply form, remove them from their point of use, isolate/quarantine the unit(s), and return. 4) Share and maintain awareness of this communication in your practice with until all required actions have been completed within your facility. 5) If you have further distributed the affected product, please notify the applicable parties about this notice. You may copy and distribute this notification letter. 6) If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. 7) If you are a distributor, note that you are responsible for notifying your affected customers. 8) Complete and return the response form via email to fieldaction@stryker.com For questions or concerns, contact fieldaction@stryker.com.
- Quantity
- not on file
- Distribution
- Worldwide - US Nationwide distribution in the states of CO, CT, FL, IL, MI, MS, NY, OH, RI, UT, VA, AL, AR, AZ, CA, IA, ID, IN, KS, KY, MA, MN, MO, MT, NC, NE, NM, NV, OK, OR, PA, TN, TX, WA, WI and the countries of Australia, Austria, Brazil, Canada, Chile, Cyprus, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Kuwait, Latvia, Lebanon, Netherlands, New Zealand, Panama, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
- Lot, serial or model codes
- Catalog/UDI-DI: CAR-06-US/00852897002328, CAR-08-US/00852897002021, CAR-10-US/ 00852897002038, CAR-12-US/00852897002335. All lots Distributed from July 2016 to October 2024.
- 510(k) numbers
- not on file
- PMA numbers
- ["P150017"]
- Product code
- PNW
- Firm FEI
- 3009351138
- Firm city
- Alpharetta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28