FDA Medical Device Recalls (openFDA)
Neotract Inc: UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
- Recall number
- Z-0598-2019
- Recalling firm
- Neotract Inc
- Product
- UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
- Status
- Terminated
- Initiated
- 2018-07-06
- Posted
- not on file
- Terminated
- 2020-12-02
- Reason
- Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
- Root cause
- Finished device change control
- Action
- On July 6, 2018, the firm sent an Urgent Medical Device Recall Notification to customers via Federal Express next day Service informing them that upon implant deployment, the capsular tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of existing known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle. It is important to note that this issue impacts only the delivery device. The implant is not impacted. Any implants that have been delivered with the device are not affected. The notification also provides instructions and the following actions to take: 1) If you HAVE affected stock: a. Immediately discontinue use and quarantine any products with the lot numbers listed above, so that the affected products can be returned to the firm. b. Please return the product as follows: i. Complete the enclosed Recall Acknowledgement Form and email to uroliftrecall@teleflex.com or fax it to (925) 401-0699, Attn: Customer Service. This will allow us to document the amount of product you have on hand for return. A customer service representative will issue you a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to the firm. ii. Write the RGA Number on each box and follow the shipping instructions provided by the firm's Customer Service. 2) If you HAVE NO affected stock: a. Please complete the enclosed Recall Acknowledgement Form and email to uroliftrecall@teleflex.com or fax it to (925) 401-0699, Attn: Customer Service. b. This will allow us to document your receipt of this letter. If customers have any other questions, they can contact their local Urology Consu
- Quantity
- 8,804
- Distribution
- US: AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV,
- Lot, serial or model codes
- Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132
- 510(k) numbers
- ["K173087"]
- PMA numbers
- not on file
- Product code
- PEW
- Firm FEI
- 3005791775
- Firm city
- Pleasanton
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28