FDA Medical Device Recalls (openFDA)
Philips Ultrasound, Inc.: Philips Lumify Diagnostic Ultrasound, Catalogue Number: 795216 Part Number: 989605449841 with 453561845331 (software version 1.0)
- Recall number
- Z-0596-2016
- Recalling firm
- Philips Ultrasound, Inc.
- Product
- Philips Lumify Diagnostic Ultrasound, Catalogue Number: 795216 Part Number: 989605449841 with 453561845331 (software version 1.0)
- Status
- Terminated
- Initiated
- 2015-12-09
- Posted
- 2016-01-06
- Terminated
- 2016-08-08
- Reason
- Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color Bar conveys that Red is to be displayed for color flow toward the Transducer, the system displays Blue for color flow toward the Transducer. This could result in misdiagnosis in some studies.
- Root cause
- Software design
- Action
- Philips began mailing the URGENT - Medical Device Correction Philips Model Lumify Ultrasound System, FSN 79500372, dated 2015 DEC 11, to consignees. Consignees are instructed to do the following: 1. Check the version of software on your system using the instructions provided above on How to Identify Affected Products. Philips has corrected this issue in version 1.0.1 and higher. 2. Connect the tablet to the internet. 3. The system will alert you to download the upgrade from Google Play. 4. Follow all on screen to upgrade the software version. If consignees cannot promptly connect to the internet to download the upgraded software, Philips recommends that you do not use Color Flow. In accordance with the subscription terms and conditions, you will be required to reconnect the system to the internet within no more than 30 days. If consignees do not connect their device to the internet after 30 days, they will be unable to use the system. Consignees with questions should contact your local Philips representative or Philips Customer Service at 1-800-722-9377.
- Quantity
- 21 units in the US only
- Distribution
- distributed in CA, CT, ND, NV, OR, RI, TN, and WA.
- Lot, serial or model codes
- System Serial numbers: USD15L0177; USD15L0178; USD15L0179; USD15L0297; USD15L0298; USN15L0789; USN15L0862; USN15L0863; USN15L0864; USN15L0865; USN15L0866; USN15L0867; USN15L0917; USN15L0918; USN15L0988; USO15L0563; USO15L0564; USO15L0565; USO15L0566; USO15L1087; and USO15L1088.
- 510(k) numbers
- ["K152899"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 3019216
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28