Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems, Inc. Cardiac & Monitoring Systems: Intellivue Information Center Model: M3150B#C01

Recall number
Z-0596-03
Recalling firm
Philips Medical Systems, Inc. Cardiac & Monitoring Systems
Product
Intellivue Information Center Model: M3150B#C01
Status
Terminated
Initiated
2003-01-02
Posted
2003-02-28
Terminated
2006-02-15
Reason
Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display, keyboard or mouse lock up
Root cause
Other
Action
Philips Medical Systems notified users by Certified Mail on/about 2/19/03 that the firm was initiating a device field correction (recall) to conduct an upgrade to units witha HP VL400 or VL420 computer which is cconfigured for dual display by replacing the video and network communication card. Previous to this notification, Philips sent letters on 1/2/03 addressed to Biomedical Engineering Manager to advise users to reboot the system in the event of a system freeze.
Quantity
39 units
Distribution
Nationwide Canada, Europe, Asia, Latin America
Lot, serial or model codes
Serial 4225A00106 4225A00107 4225A00108 4225A00111 4225A00112 4225A00113 4225A00114 4225A00116 4225A00117 4225A00118 4225A00119 4225A00127 4225A00145 4225A00149 4225A00150 4225A00151 4225A00152 4225A00164 4225A00165 4225A00166 4225A00167 4225A00169 4225A00188 4225A00189 4213A13698 4213A13700 4225A00122 4225A00136 4225A00138 4225A00137 4225A00139 4225A00140 4225A00142 4225A00143 4225A00144 4225A00101 4225A00102 4225A00103 4225A00104
510(k) numbers
["K011093"]
PMA numbers
not on file
Product code
not on file
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register