Zimmer Biomet, Inc.: Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item Number: 475665
- Recall number
- Z-0595-2020
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item Number: 475665
- Status
- Terminated
- Initiated
- 2019-10-25
- Posted
- not on file
- Terminated
- 2020-04-30
- Reason
- Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam
- Root cause
- Error in labeling
- Action
- Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL letter via FedEx and email on 10/25/19. The letter identifies the reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. Questions call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, your questions may be emailed to corporateQuality.PostMarket@zimmerbiomet.com.
- Quantity
- 34 units
- Distribution
- Distribution to US states of CA, FL, IN, MN, MO, NY, and UT, Australia, and New Zealand
- Lot, serial or model codes
- Lot Numbers: 334360, 366680, 366670 UDI: (01)00880304386587(17)280805(10)334360 (01)00880304386587(17)280530(10)366680 (01)00880304386587(17)280530(10)366670
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28