Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic MiniMed, Inc.: CareLink Clinic, REF: MMT-7350

Recall number
Z-0594-2026
Recalling firm
Medtronic MiniMed, Inc.
Product
CareLink Clinic, REF: MMT-7350
Status
Open, Classified
Initiated
2025-10-21
Posted
2025-11-25
Terminated
not on file
Reason
Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.
Root cause
Software change control
Action
On 10/21/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were emailed to customers informing them of the following: software version 4.2C was released to correct the issues with the data displayed on 24-hour Sensor Glucose Overview graph. If you made individual diabetes treatment recommendations to patients based on the hypoglycemic patterns or on the 30-day sensor glucose data displayed on the 24-hour Sensor Glucose Overview graph with affected software, revisit those recommendations utilizing patients individual CareLink reports, as they were not impacted by this issue. Complete and return the confirmation form: https://info.medtronicdiabetes.com/carelink-clinic-dashboard-hcp or email form to rs.safetyinformation@medtronic.com For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option 9.
Quantity
12126
Distribution
Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)
Lot, serial or model codes
UDI-DI: 0763000B00008748K. Software Versions: 4.2B Software Available: US: Between September 4th, 2025, and October 3rd, 2025. OUS: Between September 11th, 2025, and October 4th, 2025. US (Clinical): Between August 21st, 2025, and October 3rd, 2025. OUS (Clinical): Between August 21st, 2025, and October 4th, 2025.
510(k) numbers
not on file
PMA numbers
not on file
Product code
PHV
Firm FEI
3003166194
Firm city
Northridge
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register