Skip to the content

FDA Medical Device Recalls (openFDA)

Asante Solutions, Inc.: Asante Comfort Infusion Set with 60cm tubing & 13mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5123-4) Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark

Recall number
Z-0594-2014
Recalling firm
Asante Solutions, Inc.
Product
Asante Comfort Infusion Set with 60cm tubing & 13mm cannula packaged in 5-packs, singles,or inside a Combination Kit (FG-5123-4) Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark
Status
Terminated
Initiated
2013-11-19
Posted
2013-12-27
Terminated
2014-04-24
Reason
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Root cause
Device Design
Action
Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.
Quantity
834 kits
Distribution
Nationwide Distribution.
Lot, serial or model codes
Asante Catalog number: FG-4123-5 FG-4123T Unomedical Item number: 87-060-2562; Lot numbers: 0230118 0230122 5036994
510(k) numbers
["K120872"]
PMA numbers
not on file
Product code
FPA
Firm FEI
3009351200
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register