FDA Medical Device Recalls (openFDA)
Draeger, Inc.: ErgoStar CM 60, Model/Catalog Number: MP01860, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.
- Recall number
- Z-0593-2026
- Recalling firm
- Draeger, Inc.
- Product
- ErgoStar CM 60, Model/Catalog Number: MP01860, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.
- Status
- Open, Classified
- Initiated
- 2025-11-10
- Posted
- 2025-12-08
- Terminated
- not on file
- Reason
- Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.
- Root cause
- Under Investigation by firm
- Action
- On November 10, 2025, Urgent Medical Device Recall letters were sent to customers. Necessary Action: Identify if you have any affected product. Do not use any unused ErgoStar CM 40/CM 45/CM 55/CM 60 in your stock. Unused stock should be removed and returned to Draeger. Draeger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement. Please ensure that all potential users in your facility are made aware of this Urgent Medical Device Recall Notice and the information contained herein. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification.
- Quantity
- 2050 units
- Distribution
- Worldwide distribution. US states of CA, FL, IL, NY, OH, OK, PA, SC, TX, and WI; Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hungary, Indonesia, Italy, Japan, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkiye, Ukraine, and United Kingdom.
- Lot, serial or model codes
- Model/Catalog Number: MP01860; UDI Number: 04048675420699; Lot numbers: All lot numbers;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BZA
- Firm FEI
- 2517967
- Firm city
- Telford
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28