Philips Medical Systems, Inc.: Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician

Recall number
Z-0593-2014
Recalling firm
Philips Medical Systems, Inc.
Product
Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Status
Terminated
Initiated
2013-06-07
Posted
2013-12-27
Terminated
2017-03-22
Reason
Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10
Root cause
Labeling Change Control
Action
Philips sent a Customer Information Letter dated June 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. During an EI of the firm, the firm had not reported the correction to FDA. The Customer Information letter informs the customers what the behavior is and under what circumstances. The correction consisted of a replacement label that was provided free of charge to all units affected by this issue. Philips Healthcare representatives have been contact customers with affected devices to arrange for installation of the replacement label. To correct this issue, Philips will replace the label free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the replacement label. This issue does not affect current product shipments. I f you need any further information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377. Philips apologizes for any inconvenience caused by this problem.
Quantity
93 units
Distribution
Worldwide Distribution - US Distribution including the states of IN, MS, TX and VA., and the countries of DENMARK, FRANCE, GERMANY, SPAIN and SWITZERLAND.
Lot, serial or model codes
Serial numbers: US00558442 - US00558461 US00561519 US00558531 - US00558532 US00561578 US00558615 US00562454 US00558758 - US00558759 US00562504 - US00562527 US00560165 - US00560166 US00562640 - US00562641 US00560738 - US00560739 US00562832 US00560740 - US00560745 US00563125 US00560872 US00563231 - US00563243 US00560909 US00564032 US00561068 US00564177 US00561209 US00564409 US00561447 - US00561452 US00564793
510(k) numbers
["K051134"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register