Skip to the content

FDA Medical Device Recalls (openFDA)

Symbiosis Corp.: Product is a ChoICE 182 cm Guide Wire, intended to facilitate placement of balloon dilatation catheters and other therapeutic devices during Percutaneous Transluminal Coronary Angioplasty (PTCA) procedures. The ChoICE guide wire contains a one-piece stainless steel core, which extends throughout the length of the wire and provides steerability and torqueability. The wire is tapered and formed at the distal tip. A radiopaque platinum spring coil surrounds the distal tip of the core wire. The ChoICE Guide Wire is a single use, sterile (EO), device packaged in a carrier tube, which is held in a coil by plastic clips. The carrier tube assembly of the guide wire is sealed into a pouch. One side of the pouch is clear polyethylene/polyester laminate and the reverse side is Tyvek. The pouch assemblies are packaged in a five-pack product box. A flushing tool is included in the package to facilitate hydration of the coating with heparanized saline before use. 1, page 1, is a copy of the pouch label. Exhibit 1, page 2 is a copy of the five-pack product box label. Exhibit 2 is a copy of the Directions for Use.

Recall number
Z-0592-03
Recalling firm
Symbiosis Corp.
Product
Product is a ChoICE 182 cm Guide Wire, intended to facilitate placement of balloon dilatation catheters and other therapeutic devices during Percutaneous Transluminal Coronary Angioplasty (PTCA) procedures. The ChoICE guide wire contains a one-piece stainless steel core, which extends throughout the length of the wire and provides steerability and torqueability. The wire is tapered and formed at the distal tip. A radiopaque platinum spring coil surrounds the distal tip of the core wire. The ChoICE Guide Wire is a single use, sterile (EO), device packaged in a carrier tube, which is held in a coil by plastic clips. The carrier tube assembly of the guide wire is sealed into a pouch. One side of the pouch is clear polyethylene/polyester laminate and the reverse side is Tyvek. The pouch assemblies are packaged in a five-pack product box. A flushing tool is included in the package to facilitate hydration of the coating with heparanized saline before use. 1, page 1, is a copy of the pouch label. Exhibit 1, page 2 is a copy of the five-pack product box label. Exhibit 2 is a copy of the Directions for Use.
Status
Terminated
Initiated
2002-11-19
Posted
2003-02-27
Terminated
2008-04-15
Reason
Product is mislabled as a ChoICE Floppy Guide Wire when it is actually a ChoICE ES(Extra Support) Guide Wire.
Root cause
Other
Action
Recall Letters issued 11/19/2002 via Federal Express to 7 hospital accounts. Consignees were requested to fill out a response form and return product to their Quincy, MA facility.
Quantity
19/5-pack units
Distribution
Product was distributed to hospitals in FL, MI, GA, OR, ID AND CT. 05.
Lot, serial or model codes
Individual pouch codes are coded catalog # 12132-01, batch 4785750 expiration date 2004-05 and the five pack outer boxes are coded: Catalog No. 12132-01, batch 4806821 expiration date 2004-5.
510(k) numbers
["K962572"]
PMA numbers
not on file
Product code
DQX
Firm FEI
1051710
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register