Symbiosis Corp.: Product is a ChoICE 182 cm Guide Wire, intended to facilitate placement of balloon dilatation catheters and other therapeutic devices during Percutaneous Transluminal Coronary Angioplasty (PTCA) procedures. The ChoICE guide wire contains a one-piece stainless steel core, which extends throughout the length of the wire and provides steerability and torqueability. The wire is tapered and formed at the distal tip. A radiopaque platinum spring coil surrounds the distal tip of the core wire. The ChoICE Guide Wire is a single use, sterile (EO), device packaged in a carrier tube, which is held in a coil by plastic clips. The carrier tube assembly of the guide wire is sealed into a pouch. One side of the pouch is clear polyethylene/polyester laminate and the reverse side is Tyvek. The pouch assemblies are packaged in a five-pack product box. A flushing tool is included in the package to facilitate hydration of the coating with heparanized saline before use. 1, page 1, is a copy of the pouch label. Exhibit 1, page 2 is a copy of the five-pack product box label. Exhibit 2 is a copy of the Directions for Use.
- Recall number
- Z-0592-03
- Recalling firm
- Symbiosis Corp.
- Product
- Product is a ChoICE 182 cm Guide Wire, intended to facilitate placement of balloon dilatation catheters and other therapeutic devices during Percutaneous Transluminal Coronary Angioplasty (PTCA) procedures. The ChoICE guide wire contains a one-piece stainless steel core, which extends throughout the length of the wire and provides steerability and torqueability. The wire is tapered and formed at the distal tip. A radiopaque platinum spring coil surrounds the distal tip of the core wire. The ChoICE Guide Wire is a single use, sterile (EO), device packaged in a carrier tube, which is held in a coil by plastic clips. The carrier tube assembly of the guide wire is sealed into a pouch. One side of the pouch is clear polyethylene/polyester laminate and the reverse side is Tyvek. The pouch assemblies are packaged in a five-pack product box. A flushing tool is included in the package to facilitate hydration of the coating with heparanized saline before use. 1, page 1, is a copy of the pouch label. Exhibit 1, page 2 is a copy of the five-pack product box label. Exhibit 2 is a copy of the Directions for Use.
- Status
- Terminated
- Initiated
- 2002-11-19
- Posted
- 2003-02-27
- Terminated
- 2008-04-15
- Reason
- Product is mislabled as a ChoICE Floppy Guide Wire when it is actually a ChoICE ES(Extra Support) Guide Wire.
- Root cause
- Other
- Action
- Recall Letters issued 11/19/2002 via Federal Express to 7 hospital accounts. Consignees were requested to fill out a response form and return product to their Quincy, MA facility.
- Quantity
- 19/5-pack units
- Distribution
- Product was distributed to hospitals in FL, MI, GA, OR, ID AND CT. 05.
- Lot, serial or model codes
- Individual pouch codes are coded catalog # 12132-01, batch 4785750 expiration date 2004-05 and the five pack outer boxes are coded: Catalog No. 12132-01, batch 4806821 expiration date 2004-5.
- 510(k) numbers
- ["K962572"]
- PMA numbers
- not on file
- Product code
- DQX
- Firm FEI
- 1051710
- Firm city
- Miami
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28