Berchtold Corp.: CHROMOPHARE Surgical Lighting System & CHROMOVIEW Flat Panel Monitor, Mechanical Spring Arms.

Recall number
Z-0591-2011
Recalling firm
Berchtold Corp.
Product
CHROMOPHARE Surgical Lighting System & CHROMOVIEW Flat Panel Monitor, Mechanical Spring Arms.
Status
Terminated
Initiated
2009-11-16
Posted
2010-12-10
Terminated
2012-07-06
Reason
A mechanical failure of a welded joint within the spring Arm may result in the rapid drop of a monitor to the mechanical stop.
Root cause
Other
Action
Berchtold issued a Product Safety Notification letter on November 16, 2009 identifying the affected device and issue prompting the action. Consignees were instructed to: Exercise caution when adjusting the position of the surgical lights or flat panel monitors. While an event is highly unlikely, the likelihood does increase when the spring arm is being repositioned. Identify a specific contact person for their facility to facilitate the communication and field action process with BERCHTOLD. The following action is being planned: To further identify if their units are part of the affected population, determine the supplier serial number identification and specific location for each spring arm in their facility. Report the facility information, supplier spring arm serial numbers and specific location or spring arms to a BERCHTOLD. Detail instructions on serial number identification and the method of reporting to BERCHTOLD will also be provided in a subsequent communication. Once BERCHTOLD has received the serial numbers of their spring arm assembly, customers will be notified whether their particular units are affected and the further required action. Berchtold Corporation can be contacted at 1 800 968 6097 about this action.
Quantity
7031 arms
Distribution
Worldwide Distribution: Throught the U.S,.and the countries of Canada and Germany
Lot, serial or model codes
Spring Arm Serial Numbers: 010103837915-38114, 010101835874-35973, 010101836138, 010103837680-37779, 010101836309-36408, 10109841913-41942, 010107841041-41060, 010106840378-40407, 010104838741-38770, 010102837580-37609, 010102837341-37370, 01010183669, 010112735745-35804, 010112735193-35302, 010111734501-34663, 010103838406-38555, 010101835988-36137, 010103838574-38603, 010102837465-37494, 010102837301-37340, 010112735125-35134, 010109946662-46711, 010106945840-45866, 010110842349-42388, 010107841061-41100, 010107840798-40837, 010107840778-40797, 010103838286-38325, 010103838286-38325, 010102836829-36868, 010109946642-46661, 010107946042-46050, 010109842141-42180, 010109842140-42140, 010109841871-41910, 010108841499-41538, 010107840959-40998, 010106840715-40754, 010104839108-39147, 010105839565-39604, 010104839098-39107, 010104838907-38961, 010102837525-37578, 010102837579, 010101836778-36813, 010101835974-35977, 010204803835-03844, 010202802298-02307, 010202802223-02232, 010201801889-01898, 010201801569-01578, 010211807820-07969, 010211807720-07819, 010210807565-07614, 010210807454-07553, 010210807254-07453, 010209806644-06893, 010209806504-06603, 010208806352-06501, 010208806262-06311, 010208806092-06141, 010207805722-05771, 010207805479-05681, 010206805229-05328, 010206805079-05228, 010205804650-04749, 010205804358-04407, 010204803965-04164, 010204803965-04014, 010204803535-03634, 010202802579-02628, 010201802039-02088, 010201801839-01888, 010201801189-01338, 010205804500-04649, 010204804220-04269, 010205804408-04457, 010204804170-04219, 010204804065-04164, 010204803635-03734, 010203803485-03534, 010203803375-03424, 010203803275-03374, 010203802709-02808, 010202802308-02407, 010202802408-02507, 010201801899-01998, 010201801639-01738, 010201801739-01838, 010201801089-01188, 010212808270-08409, 010211808210-08269, 010209806994-07053, 010209806894-06993, 010209806604-06643, 010208806312-06351, 010208806202-06261, 010208806142-06201, 010207805979-06038, 010207805772-05871, 010204803865-03964, 010104803735-03834, 010203802909-03008, 010201802089-02188, 010201801999-02038, 010201801579-01638, 010207805892-05911, 010207805872-05891 and 010105839563-39564.
510(k) numbers
["K922836"]
PMA numbers
not on file
Product code
FTD
Firm FEI
1220685
Firm city
N Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register