Skip to the content

FDA Medical Device Recalls (openFDA)

Bausch & Lomb Inc: Bausch & Lomb SoFlex SE Foldable Intraocular Lens. UPC 4001404, Rx Only, STERILE. Manufactured by: Bausch & Lomb Incorporated, 21 Park Place Boulevard, Clearwater, FL 33759. Intraocular lens indicated for primary implantation for visual correction in adult patients where the cataractous lens has been removed by phacoemulsification. The lens is intended for placement in the capsular bag.

Recall number
Z-0591-2010
Recalling firm
Bausch & Lomb Inc
Product
Bausch & Lomb SoFlex SE Foldable Intraocular Lens. UPC 4001404, Rx Only, STERILE. Manufactured by: Bausch & Lomb Incorporated, 21 Park Place Boulevard, Clearwater, FL 33759. Intraocular lens indicated for primary implantation for visual correction in adult patients where the cataractous lens has been removed by phacoemulsification. The lens is intended for placement in the capsular bag.
Status
Terminated
Initiated
2009-11-19
Posted
2010-01-11
Terminated
2010-07-22
Reason
Bausch & Lomb Incorporated, Rochester, NY is recalling their Silicone L161SE Intraocular Lens Lot 4916928 (23.5D) after receiving three customer complaints for what appears to be opacification of the intraocular lens. The opacification occurred within one week of implantation. In each reported case the surgeon explanted the lens.
Root cause
Other
Action
Bausch & Lomb, Inc. notified consignees via phone beginning November 19, 2009 and by follow-up letter dated November 23, 2009. Titled "Urgent - Medical Device Recall", the letter instructed users to discontinue implantation of the affected product and return any unused lenses along with the enclosed acknowledgement form to the firm. For further information, contact Bausch & Lomb at 1-585-338-6612.
Quantity
85 lenses
Distribution
Lenses were distributed to one distributor and 35 retailers. Worldwide Distribution -- Distributed to Canada, Taiwan and throughout the United States.
Lot, serial or model codes
Model Number: LI61SE, Lot Number: 4916928.
510(k) numbers
not on file
PMA numbers
["P910061"]
Product code
HQL
Firm FEI
1317628
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register