Skip to the content

FDA Medical Device Recalls (openFDA)

Hocoma AG: Hocoma Lokomat-System with Pediatric Module Software Versions: V4.30, V4.31, V4.31a, V4.31b, V4.31c, V4.32, V4.32a, V5.0x. Powered exercise equipment used to apply automated locomoter training for hemiplegic and paraplegic patients.

Recall number
Z-0590-2010
Recalling firm
Hocoma AG
Product
Hocoma Lokomat-System with Pediatric Module Software Versions: V4.30, V4.31, V4.31a, V4.31b, V4.31c, V4.32, V4.32a, V5.0x. Powered exercise equipment used to apply automated locomoter training for hemiplegic and paraplegic patients.
Status
Terminated
Initiated
2009-05-27
Posted
2010-01-07
Terminated
2010-02-22
Reason
Changes to the stored "sensitivity" values may occur and leads to higher limits in the force and trajectory deviation
Root cause
Software design
Action
Hocoma issued a letter dated May 27, 2009 to Lokomat users informing them of the affected device. Instructions were provided to avoid further problems with the product and users were asked to complete and return the provided Declaration to Field Safety Notice. For further information, contact Hocoma at info@hocoma.com.
Quantity
7
Distribution
Nationwide Distribution -- including IL, MA, MI, NY and OH.
Lot, serial or model codes
Software Versions: V4.30, V4.31, V4.31a, V4.31b, V4.31c, V4.32, V4.32a and V5.0x.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BXB
Firm FEI
3003781275
Firm city
Zuerich
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register