Cardiac Science Corporation: GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.

Recall number
Z-0589-2011
Recalling firm
Cardiac Science Corporation
Product
GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.
Status
Terminated
Initiated
2010-04-28
Posted
2010-12-10
Terminated
2011-12-08
Reason
Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.
Root cause
Nonconforming Material/Component
Action
Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.
Quantity
130 units
Distribution
Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.
Lot, serial or model codes
Serial Numbers:  3201281, 3201806, 3201810, 3201815, 3202223, 3202233, 3202244, 3202248, 3202249, 3202253, 3202255, 3202256, 3202257, 3202260, 3202262, 3202269, 3202271, 3202273, 3202275, 3202280, 3202281, 3202283, 3202284, 3202285, 3202286, 3202287, 3202288, 3202290, 3202291, 3202292, 3202297, 3202299, 3202300, 3202301, 3202303, 3202304, 3202305, 3202306, 3202308, 3202310, 3202311, 3202312, 3202313, 3202315, 3202316, 3202317, 3202319, 3202321, 3202324, 3202327, 3202328, 3202331, 3202332, 3202333, 3202334, 3202336, 3202337, 3202338, 3202339, 3202340, 3202344, 3202347, 3202348, 3202349, 3202350, 3202351, 3202352, 3202354, 3202355, 3202356, 3202357, 3202358, 3202360, 3202361, 3202362, 3202363, 3202364, 3202365, 3202371, 3202377, 3202379, 3202380, 3202381, 3202382, 3202385, 3202386, 3202387, 3202388, 3202389, 3202390, 3202391, 3202392, 3202393, 3202394, 3202395, 3202396, 3202397, 3202398, 3202399, 3202400, 3202401, 3202403, 3202404, 3202405, 3202406, 3202407, 3202408, 3202409, 3202411, 3202412, 3202414, 3202415, 3202416, 3202417, 3202418, 3202419, 3202420, 3202421, 3202422, 3202423, 3202427, 3202429, 3202439, 3202442, 3202443, 3202444, 3202447, 3202448, 3202450, and 3202451.
510(k) numbers
["K060934"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
2112020
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register