Cardiac Science Corporation: GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.
- Recall number
- Z-0589-2011
- Recalling firm
- Cardiac Science Corporation
- Product
- GE Responder 2000 Defibrillator/Monitor, models 2025653, 2026109, and 2026114 that were manufactured between July 25, 2008 and October 2, 2009 by GE Healthcare.
- Status
- Terminated
- Initiated
- 2010-04-28
- Posted
- 2010-12-10
- Terminated
- 2011-12-08
- Reason
- Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.
- Root cause
- Nonconforming Material/Component
- Action
- Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.
- Quantity
- 130 units
- Distribution
- Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.
- Lot, serial or model codes
- Serial Numbers: 3201281, 3201806, 3201810, 3201815, 3202223, 3202233, 3202244, 3202248, 3202249, 3202253, 3202255, 3202256, 3202257, 3202260, 3202262, 3202269, 3202271, 3202273, 3202275, 3202280, 3202281, 3202283, 3202284, 3202285, 3202286, 3202287, 3202288, 3202290, 3202291, 3202292, 3202297, 3202299, 3202300, 3202301, 3202303, 3202304, 3202305, 3202306, 3202308, 3202310, 3202311, 3202312, 3202313, 3202315, 3202316, 3202317, 3202319, 3202321, 3202324, 3202327, 3202328, 3202331, 3202332, 3202333, 3202334, 3202336, 3202337, 3202338, 3202339, 3202340, 3202344, 3202347, 3202348, 3202349, 3202350, 3202351, 3202352, 3202354, 3202355, 3202356, 3202357, 3202358, 3202360, 3202361, 3202362, 3202363, 3202364, 3202365, 3202371, 3202377, 3202379, 3202380, 3202381, 3202382, 3202385, 3202386, 3202387, 3202388, 3202389, 3202390, 3202391, 3202392, 3202393, 3202394, 3202395, 3202396, 3202397, 3202398, 3202399, 3202400, 3202401, 3202403, 3202404, 3202405, 3202406, 3202407, 3202408, 3202409, 3202411, 3202412, 3202414, 3202415, 3202416, 3202417, 3202418, 3202419, 3202420, 3202421, 3202422, 3202423, 3202427, 3202429, 3202439, 3202442, 3202443, 3202444, 3202447, 3202448, 3202450, and 3202451.
- 510(k) numbers
- ["K060934"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 2112020
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28