Burton Medical Products Corp: All Outpatient and Coolspot single ceiling or single track Medical Examination Lights manufactured before August 2000. Products manufactured during or after August 2000 contain a redesigned pivot assembly and are not included in this recall. Affected products include: Model 100540 Fleximount SGL Ceiling, Serial No.''s less than SC995486 Model 100740 SGL Ceiling CS II, Serial No''s less than SCCS990745 Model 100580 SGL Trolley, Serial No.''s less than TRS962633 Model 100780 SGL Trolley OP II, Serial No.''s less than STOP97417 Model 102540 SGL Ceiling OP II, Serial No.''s less than SCOP003354.

Recall number
Z-0589-04
Recalling firm
Burton Medical Products Corp
Product
All Outpatient and Coolspot single ceiling or single track Medical Examination Lights manufactured before August 2000. Products manufactured during or after August 2000 contain a redesigned pivot assembly and are not included in this recall. Affected products include: Model 100540 Fleximount SGL Ceiling, Serial No.''s less than SC995486 Model 100740 SGL Ceiling CS II, Serial No''s less than SCCS990745 Model 100580 SGL Trolley, Serial No.''s less than TRS962633 Model 100780 SGL Trolley OP II, Serial No.''s less than STOP97417 Model 102540 SGL Ceiling OP II, Serial No.''s less than SCOP003354.
Status
Terminated
Initiated
2004-02-19
Posted
2004-07-20
Terminated
2012-01-25
Reason
Pivot joint on these ceiling mount lights may fail and lights may fall onto patients.
Root cause
Component design/selection
Action
Firm will ask for subdistribution information from thier distributors via letter on 1/30/2004. Letters will be sent directly to users. Firm will retrofit all sites with suspect units with new pivot assemblies.
Quantity
17,000 estimated
Distribution
Nationwide.
Lot, serial or model codes
See Product Description
510(k) numbers
["K851742"]
PMA numbers
not on file
Product code
KZF
Firm FEI
2018492
Firm city
Chatsworth
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register