Skip to the content

FDA Medical Device Recalls (openFDA)

Thornhill Research Inc: MOVES System, an emergency ventilator with suction, oxygen concentrator and multi-parameter patient monitoring capabilities. The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.

Recall number
Z-0586-2016
Recalling firm
Thornhill Research Inc
Product
MOVES System, an emergency ventilator with suction, oxygen concentrator and multi-parameter patient monitoring capabilities. The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.
Status
Terminated
Initiated
2015-11-19
Posted
2016-01-11
Terminated
2019-04-12
Reason
Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.
Root cause
Component design/selection
Action
Thornhill Research sent a Recall Notification letter dated December 15, 2015 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to update their device software. For questions customers were instructed to call 416-597-1325.
Quantity
83 units
Distribution
Worldwide Distribution: US distribution in state of: Virginia and to the country of: Singapore.
Lot, serial or model codes
MS201205001, MS201205002, MS201205003, MS201205004, MS201205005, MS201212001, MS201212002, MS201212003, MS201212004, MS201212005, MS201212006, MS201212007, MS201212008, MS201212009, MS201212010, MS201212011, MS201212012, MS201212013, MS201212024, MS201212025, MS201212026, MS201212028, MS201212029, MS201212030, MS201212031, MS201212032, MS201212033, MS201212034, MS201212035, MS201212036, MS201212037, MS201212038, MS201212039, MS201212040, MS201212041, MS201212042, MS201212053, MS201212054, MS201212055, MS201212056, MS201212057, MS201212058, MS201212059, MS201212060, MS201212061, MS201212062, MS201212062, MS201212063, MS201212064, MS201212065, MS201212066, MS201212067, MS201212068, MS201212069, MS201212070, MS201212014, MS201212015, MS201212016, MS201212017, MS201212018, MS201212019, MS201212020, MS201212021, MS201212022, MS201212023, MS201212043, MS201212044, MS201212045, MS201212046, MS201212047, MS201212048, MS201212049, MS201212050, MS201212051, MS201212052, MS201212071, MS201212072, MS201212073, MS201212074, MS201212075, MS201212076, MS201212077, MS201212078, MS201212079
510(k) numbers
["K093261"]
PMA numbers
not on file
Product code
BTL
Firm FEI
3005803793
Firm city
Toronto
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register