FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: TIBIO FEM/FEM NAIL.
- Recall number
- Z-0586-2014
- Recalling firm
- Zimmer, Inc.
- Product
- TIBIO FEM/FEM NAIL.
- Status
- Terminated
- Initiated
- 2013-11-05
- Posted
- 2013-12-27
- Terminated
- 2015-05-05
- Reason
- Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
- Root cause
- Process change control
- Action
- On 11/5/2013, Urgent Medical Device Recall Notifications were sent to affected distributors, hospital Risk Managers, and doctors informing them of the failure. All distributors were notified via electronic mail. Hospital risk managers, surgeons and distributors with affected inventory within the expiration date of the device will also be notified via courier. The letter identifies the issue, health risks, and their responsibilities. Questions concerning the recall are directed to 1-877-946-2761 Communications outside of the United States will occur approximately two weeks after the United States communications. Zimmer will conduct effectiveness checks in the following way: Distributors/Hospitals/Surgeons 100% of the notifications will be tracked to ensure delivery of the notifications. An inventory certification form will need to be returned from each distributor that certifies that all available affected product from their territory has been returned. Additional notifications via email and FedEx will be sent to those who have not completed the required certification form or returned affected product. Accounts will be deemed unresponsive after 3 attempts.
- Quantity
- 42, 065 units total
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Item Numbers 32-8553-311-11 32-8553-311-13 32-8553-311-15 32-8553-311-17 All lots with an expiration date prior to July 24, 2023
- 510(k) numbers
- ["K130595"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28