Skip to the content

FDA Medical Device Recalls (openFDA)

B. Braun Medical Inc: B. Braun Hyperformer EVA Mixing Container with Hyperformer Compatible Connector, 4000ml size, Catalog Number SH4000, Batch Numbers 05A14474, 05A18561, and 05A19926

Recall number
Z-0586-06
Recalling firm
B. Braun Medical Inc
Product
B. Braun Hyperformer EVA Mixing Container with Hyperformer Compatible Connector, 4000ml size, Catalog Number SH4000, Batch Numbers 05A14474, 05A18561, and 05A19926
Status
Terminated
Initiated
2006-02-06
Posted
2006-03-03
Terminated
2008-05-28
Reason
B. Braun Medical has been notified by the manufacturer of the product (Stedim) that a small percentage of the lots leak from the set port when the set port is opened. The leak was identified by Stedim to be due to the incorrect placement of a septum in the set port. The set closure is intact and the leak remains internal until the set port is opened.
Root cause
Other
Action
A recall notification letter was sent to the 22 customers who received the product between 12/1/05 and 1/11/06. The recall packages included a letter addressed to either the Director of Pharmacy/Director of Nursing, Materials Manager, or Distributor, as appropriate. They were mailed via US Post Office certified return receipt mail. Customers were instructed by letter to return all product at their facility to B. Braun Medical, Inc. in Breinigsville Pennsylvania. A Recall Information Form and a postage paid B. Braun Medical, Inc. return envelope were included in the recall package for the customer''s use.
Quantity
13,480
Distribution
Nationwide to: CA, CT, FL, GA, KS, MD, MI, NC, NJ, NY, SC, TN, & TX
Lot, serial or model codes
Lot Numbers: 05A14474, 05A18561, & 05A19926
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPE
Firm FEI
2021236
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.