Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Exactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-35-3511, h) 02-022-35-4511, i) 02-022-44-3512, j) 02-022-47-2509, k) 02-022-51-4519

Recall number
Z-0585-2024
Recalling firm
Exactech, Inc.
Product
Exactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-35-3511, h) 02-022-35-4511, i) 02-022-44-3512, j) 02-022-47-2509, k) 02-022-51-4519
Status
Open, Classified
Initiated
2023-11-21
Posted
2024-01-18
Terminated
not on file
Reason
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
Root cause
Packaging change control
Action
Exactech issued an URGENT MEDICAL DEVICE RECALL notice to its direct consignees on 11/21/2023 via email. The notice explained the issue, clinical impact, and requested the removal and return of all affected products. 11/22/2023, the firm initiated the recall OUS also by email.
Quantity
not on file
Distribution
US NY, TX, IN, IL, MN, KY, VA, MD, TN, GA, NH, ME, SC, NV, FL, AR and Australia, Brazil, Chile, France, Germany, Guatemala, Italy, Japan, Korea, Slovenia, Spain, UK
Lot, serial or model codes
a) 02-012-35-3509, GTIN 10885862304346, Serial Numbers: A730624, A730626, A730628, A730630, A730631, A730632, A730633; b) 02-022-35-1509, GTIN 10885862303974, Serial Numbers: A641578, A683756, A683757, A683758, A683759; c) 02-022-35-1513, GTIN 10885862304018, Serial Numbers: A662293; d) 02-022-35-3010, GTIN 10885862304261, Serial Numbers: A662100, A662102, A662104, A662105, A662106, A662107, A662109; e) 02-022-35-3509, GTIN 10885862304346, Serial Numbers: A652047, A652048, A652050, A652052, A652053, A652054, A730952, A730960, A730961; f) 02-022-35-3510, GTIN 10885862304353, Serial Numbers: A662121; g) 02-022-35-3511, GTIN 10885862304360, Serial Numbers: A662301, A662304, A662305; h) 02-022-35-4511, GTIN 10885862304544, Serial Numbers: A684499; i) 02-022-44-3512, GTIN 10885862305206, Serial Numbers: A641603, A641606, A641608, A641612; j) 02-022-47-2509, GTIN 10885862306326, Serial Numbers: A658478, A658486, A702263, A702265, A702267, A702268, A702269, A702271; k) 02-022-51-4519, GTIN 10885862307866, Serial Numbers: A737101, A737106, A737107;
510(k) numbers
["K152170","K171045"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register